Creatinine Reagent Lot Positive Bias – Resolution

March 18 2021
On March 3rd 2021, LifeLabs sent a memo to all clients informing them of a positive bias on the serum creatinine assay, causing a positive shift of up to 8 umol/L in creatinine results, along with the concurrent negative shift in eGFR. The root cause was the manufacturer reagent lot bias. An interpretive comment was added to all serum creatinine results at that time, to inform clients of the shift, while LifeLabs worked on resolution. A new reagent lot has been made available recently, and LifeLabs validation has confirmed that it corrects the previously introduced bias. This lot is implemented at all LifeLabs laboratory sites as of March 18, 2021.

On March 3rd 2021, LifeLabs sent a memo to all clients informing them of a positive bias on the serum creatinine assay, causing a positive shift of up to 8 umol/L in creatinine results, along with the concurrent negative shift in eGFR. The root cause was the manufacturer reagent lot bias. An interpretive comment was added to all serum creatinine results at that time, to inform clients of the shift, while LifeLabs worked on resolution.

A new reagent lot has been made available recently, and LifeLabs validation has confirmed that it corrects the previously introduced bias. This lot is implemented at all LifeLabs laboratory sites as of March 18, 2021.

Learn more details

9

CHANGES TO REFERRED OUT TESTING LOCATIONS

As part of a large process and quality improvement initiative around our referred out testing progra...
1

LIPOPROTEIN (A)

Due to supply issues at our referral laboratory, we are currently experiencing a delay in turnaround...
2

Changes to Urinalysis Platform – September 2021

LifeLabs is upgrading our macroscopic and microscopic urinalysis instrumentation at our labs in Brit...
3

Change to ordering process for ANA & ENA tests in BC effective August 8, 2021

Previously, ANA and ENA could be ordered separately or together. Moving forward, ENA testing will no...
27

Urine Urobilinogen/Bilirubin

**Discontinuation of Tests** URINE BILIRUBIN AND UROBILINOGEN will be discontinued as of June 14, 2...
5

ANGIOTENSIN CONVERTING ENZYME

**Update on Turn Around Times** Due to supplier issues, we are currently experiencing a delay in tu...
18

Creatinine Reagent Lot Positive Bias – Resolution

On March 3rd 2021, LifeLabs sent a memo to all clients informing them of a positive bias on the seru...
18

FastTRK Testing Program

FastTRK is a clinical testing program for the diagnosis of NTRK gene fusions. Funded by Bayer Inc., ...
11

Creatinine Reagent Lot Positive Bias Update

A new creatinine reagent was started at all LifeLabs sites in BC on March 11, 2021. Creatinine and e...
5

Creatinine Reagent Lot Positive Bias.

LifeLabs has recently become aware of a positive bias on the serum creatinine assay. Test results fo...

HCPs
Newsletters

Read More
The stethoscope on the doctor's Desk

Scientific Publications

Read More
Scientist looking through a microscope in a laboratory

Your health & safety is important to us

Please note that face masks are mandatory at all times at all of our locations to ensure the health and safety of all customers and employees.

Please inform staff if you are unable to wear a mask due to a medical condition. 

Thank you for your cooperation!